高低溫試驗(yàn)箱TEMPERATURE TEST
Product Model | high and low temperature testing chamber MAX-TL720-10 |
| The equipment is prohibited to test below: Flammable, explosive, volatile substance's sample testing or storage; Test samples or storage of corrosive substances; Testing or stockpiling of biological specimens; Strong electromagnetic radiation source sample's test or storage; Testing or storage of radioactive material sample; Test samples or store strong toxic substances; |
| |
| 80~1000L,品達(dá)公司可承制非標(biāo)試驗(yàn)設(shè)備。 |
| |
| An ambient temperature of + 5 ~ + 28 ℃, relative humidity ≤85%,under conditions of no sample within the test chamber; |
| GB/T 5170.2-2008 Temperature Test Equipment GB/T 5170.5-2008 Humidity Test Equipment |
| -70℃~+150℃ |
| Temperature:±0.2℃ (The difference between the controller setpoint and actual value of the controller) |
| ≤0.5℃(Temperature fluctuation is half the difference between the center of the highest measured temperature and the minimum temperature.) |
| ≤±1℃(The average temperature of the studio temperature controller displays the value to minus the average temperature measured of the center) |
| ≤2.0℃(Temperature uniformity is each difference in the measured maximum and minimum temperatures of the test of arithmetic mean.) |
| 3℃/min(Nonlinear load) |
| 0.75~1.2℃/min(Nonlinear load) |
4.11 Working noise | A Sound Level≤70dB(A) (At ambient temperature 25 ℃ and measured in the room of echo less interior noise; Using A-weighted, average of eight test points; Each test point level from the noise source 1 m, the height from the ground 1 meter.) |
4.11. Meet the test method
| GB/T2423.1-2008(IEC60068-2-1:2007) Low Temperature Test Method Ab GB/T2423.2-2008(IEC60068-2-2:2007) High Temperature Test Method Bb GJB150.4-1986 Low temperature test GJB150.3-1986 High temperature test GB/T2423.3-2006(IEC60068-2-78:2007)Constant damp heat test method Cab GB/T2423.4-2008(IEC60068-2-30:2005) Alternating hot and humid test method Db GJB150.9-1986 Humidity Test(Picture1, Picture2) (Per cubic meter load of not more than 35kg / m3 steel thermal capacity,test when no active wet heat, heat load) |